T500
B & W Transmission Densitometer

PROVEN
PERFORMANCE & DEPENDABILITY
Ihara's
advanced micro technology ensures reliable results. The simple push-button
operation makes density reading easy.
ACCURATE DOT AREA MEASUREMENTS
Make positive or negative film dot area readings.
BROAD DENSITY RANGE
Accurate film density measurements value up to 6.00D.
COMPACT LIGHTWEIGHT DESIGN
Our cordless, portable design makes our densitometer easy to operate
at any location.
INSTANT MINI-LIGHT TABLE
The T500 lights up automatically when the measurement arm is lowered
or if any function keys are selected. Highlight target area quickly
using the mini-light table.
QUICK CHARGES
Charge battery in 1 to 1.5 hours. Orange/green LED clearly displays
charge status.
EXTEND USAGE
Each charge provides hundreds of readings. The T500 shuts off automatically
to extend battery life.
FLEXIBLE PRINTER & PC INTERFACE
RS-232 serial port links to printers for hard copy and to computers
for data analysis and storage.
Ihara T500
Black & White Transmission
Densitometer
Part Number EQP-IHT500
Black & White Reflection/Transmission Densitometer Part Number
EQP-IHTR500

Ask about our 14 day trial!
SPECIFICATIONS
|
MODEL |
Portable Black and White Transmission Densitometer T500 |
Portable Black and White Reflection Densitometer R500 |
|
DIMENSIONS |
Console: 9.9"(L) x 4.1"(W) x 3.4"(H) (251mm~103mm~87mm)
|
Console: 8.3"(L) x 2.8"(W) x 2.0"(H) (210mm~72mm~50mm)
|
|
Light Table: 3.1" x 2.8" (80mm~70mm) |
- |
|
Measuring Arm Length: 7.3" (185mm) |
- |
|
WEIGHT |
Approximately 1,300g (including battery) |
Approximately 520g (including battery) |
|
POWER SOURCE |
AC adapter DC 9V-500mA<>AC 100V(120V,220V) |
AC adapter DC 9V-500mA<>AC 100V(120V,220V) |
|
Built-in Rechargeable Battery DC4.8V-800mAh |
Built-in Rechargeable Battery DC4.8V-800mAh |
|
OPTICAL DESIGN |
ANSI PH2.19 (Specular-Diffuse) |
0/45degree (ISO 5/4, ANSI PH2.17, DIN 16536) |
|
RESPONSE |
Visual |
Visual (ISO 5/3) |
|
LIGHT SOURCE |
Halogen Lens Lamp |
Halogen Lens Lamp |
|
DETECTOR |
Filter, Built-in Photo Diode |
Filter, Built-in Photo Diode |
|
APERTURE |
ƒÓ3mm,
ƒÓ2mm, ƒÓ1mm |
3mm(standard), 1.7mm |
|
MEASURING RANGE |
Density: 0.00D - 6.00D |
Density: 0.00D - 2.50D |
|
Dot Percentage: 0% to 100% |
Dot Percentage: 0% to 100% |
|
MEASUREMENT FUNCTIONS |
Density |
Density |
|
Positive Dot Area % |
Dot Area % |
|
Negative Dot Area % |
- |
|
REPEATABILITY |
}0.01D(at<4.0D,
3mm aperture) |
}0.01D(at<2.0D),
}1% |
|
INTER-INSTRUMENT AGREEMENT |
}0.02D(at<4.0D,
3mm aperture) |
}0.02D(at<2.0D),
}2% |
|
WARM UP TIME |
None |
None |
|
FRINGE CORRECTION |
Supported |
- |
|
CALIBRATION |
One touch zeroing operation, Slope calibration using calibration
film |
Calibration using calibration reference |
|
INTERFACE |
RS-232C Serial Interface for printer (9,600bps, 8 Data
bits, Non Parity, 1 Stop bit) (serial connection port for R500)
@ |
RS-232C Serial Interface (9,600bps, 8 Data bits, Non Parity,
1 Stop bit) (serial connection port for T500) @ |
|
DISPLAY |
8 character LCD |
None |
|
OPERATIONAL KEYS |
Interlocking arm measuring switch, 2 operational panel keys
|
6 sheet keys |
|
SOUND CONTROL |
Supported |
Supported |
|
AUTO POWER OFF TIMER |
Supported |
Interlock with T500 |
|
OPERATING TEMPERATURE RANGE |
41 - 104 ‹F (5 - 40 ‹C) |
41 - 104 ‹F (5 - 40 ‹C) |
|
STANDARD ACCESSORIES |
Instruction Manual |
Instruction Manual |
|
Aperture ƒÓ3mm (installed), ƒÓ2mm, ƒÓ1mm |
Calibration Reference |
|
AC Adapter (as specified) |
Soft Case |
|
Calibration Film |
Serial Communication Cable 8P-8P |
Specifications
subject to change without notice.
Revised
February 27, 2010
|
GWJ co
Hacienda
Heights, CA 91745
Phone 626-336-1134
800-336-1138(USA)
Fax 626-336-2409
Copyright
© 1997-2010 GWJ Co. All rights reserved.
Other products and companies referred to herein are trademarks
or registered trademarks of
their respective companies.
|